Model |
SUPER05 |
Charger Input |
100–240V, Frequency: 50/60Hz, Power: 30VA |
Internal Power Supply |
14.8V |
Power Internal Resistance |
≤2Ω |
Main Unit Input Voltage |
19V– |
Radiation Type |
X-ray |
Radiation Direction & Distribution |
Collimator diameter: 60mm |
Dose Rate |
2mGy/s |
X-ray Tube Model |
XD10F-0.2/70 |
Target Material |
Tungsten |
Target Angle |
15° |
X-ray Focal Spot |
0.4mm |
Tube Voltage |
70kV ±10% |
Tube Current |
2mA ±20%; Exposure Time Adjustment Range: 0.04s–2.0s; Error: ±(10% + 1ms) |
Nominal Power |
0.14kW (70kV, 2mA, 0.1s) |
Max Output Power Load Condition |
70kV, 2mA |
Inherent Filtration |
1.0mmAl @70kV (YY/T0062-2004) |
Additional Filtration |
0.5mmAl @70kV (YY/T0062-2004) |
Total Filtration |
1.5mmAl @70kV (YY/T0062-2004) |
Half Value Layer |
≥1.6mmAl @70kV |
Duty Cycle |
1/15 |
Leakage Radiation Load Factor |
1s exposure, 15s pause; Leakage radiation rate <0.25mGy/h @1m (70kV, 2mA, 1s, 1s/15s interval) |
Weight |
2Kg |
Tube Assembly Weight |
1.1Kg |